Tektronix Telescope 1502C Bedienungsanleitung Seite 60

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 126
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 59
Operator Tutorial
210
1502C MTDR User Manual
O
F
F
O
F
F
O
F
F
O
N
ac
>=1 K W
578.000 ft
Figure 212: Ohms-at-Cursor Beyond Reflected Pulse
9. Turn the
n
o
POSITION control to move the cursor to the far left side of the
display (2.000 ft). Note that the ohms-at-cursor reading is now < 1 W.
O
F
F
O
F
F
O
F
F
O
N
ac
< 1 W
2.000 ft
Figure 213: Ohms-at-Cursor Beyond Reflected Pulse
If the cursor is placed too near a fault, the reflection will not have stabilized,
which will make the ohms-at-cursor reading misleading. This is especially true
very near the instrument where some aberrations are still significant. See the
Ohms-at-Cursor section of the Operating Instructions chapter for more on the
limitations of this feature.
Seitenansicht 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 125 126

Kommentare zu diesen Handbüchern

Keine Kommentare