Instruction ManualP5205High Voltage Differential Probe070-9472-02www.tektronix.com

Seite 2

General Safety SummaryviP5205 Instructi on ManualDo Not Operate With Suspected Failures. If you suspect there is damageto this product, ha ve it i nsp

Seite 3 - WARRANTY

P5205 Instructi on ManualviiService Safety Su mmaryOnly qualified personnel should perform service procedures. Readthis Se rvice Safety Summary and th

Seite 4

Service Safety SummaryviiiP5205 Instructi on ManualContacting TektronixPhone 1-800-833-9200*Address Tektronix, Inc.14200 SW Karl Braun DriveP.O. Box 5

Seite 5 - Table of Contents

P5205 Instructi on Manual1Getting StartedThis section describes the High Voltage Differential Probe and givesinstructions on how to install and functi

Seite 6

Getting Started2P5205 Instructi on ManualONFigure 1: High Voltage Differential Probe

Seite 7 - List of Figures

Getting StartedP5205 Instructi on Manual3The P5205 has several features and accessories that make probingand measurement a simpler task. Take a moment

Seite 8 - List of Tables

Getting Started4P5205 Instructi on ManualInput Leads. The input leads of the differential probeconnect to the crocodile clips and plunger clamps thatc

Seite 9 - General Safety Summary

Getting StartedP5205 Instructi on Manual5Zero Adjust. T he zero adjust allows you t o set t he probeoutput to the zero reference point prior to making

Seite 10 - Safety Terms and Symbols

Getting Started6P5205 Instructi on ManualInstallationInstall the P5205 probe as follows:1. Connect the output of the probe to the TEKPROBE input of th

Seite 11 - Service Safety Su mmary

Getting StartedP5205 Instructi on Manual7Functional CheckTo make a simple functional check of the P5205 probe, select asource that supplies AC line vo

Seite 12 - Contacting Tektronix

Copyright © Tektronix, Inc. All rights reserved.Tektronix products are covered by U.S. and foreign patents, issued andpending. Information in this pub

Seite 13 - Getting Started

Getting Started8P5205 Instructi on Manual

Seite 14 - P5205 Instructi on Manual

P5205 Instructi on Manual9Operating BasicsTo help you use the High Voltage Differential Probe safe ly andeffectively, this section provides important

Seite 15

Operating Basics10P5205 Instructi on ManualMinimizing Risk of RF Burn (probe leads)WARNING. To avoid personal injury, do not handle the probe leadswhe

Seite 16

Operating BasicsP5205 Instructi on Manual11Voltage(RMS)1000 V100 VDC or 0 10 M 100 M1MFrequency (Hz)5M 50M500 V50 V100 k 500 kCategory II maximumvolta

Seite 17

Operating Basics12P5205 Instructi on ManualBoth ranges are rated for t he m aximum input limits, but you canobtain a useful measurement only if the vo

Seite 18 - Installation

Operating BasicsP5205 Instructi on Manual13Figure 3: Twisting the input leadsExtension LeadsThe extension leads allow you to reach widely spaced conne

Seite 19 - Functional Check

Operating Basics14P5205 Instructi on ManualWith extension leadsWithout extension leads50 ns/div 50 ns/div250 mv/div250 mv/divFigure 4: HF Transient re

Seite 20

P5205 Instructi on Manual15Specificatio nsThe specifications in Tables 2 through 6 apply to a P5205 probeinstalled on a Tektronix TDS 460A oscilloscop

Seite 21 - Operating Basics

Specifications16P5205 Instructi on ManualTable 2: Warranted Electrical Characteristics (Cont.)Maximum Operating Input Voltage 500X differential: ± 1.3

Seite 22

SpecificationsP5205 Instructi on Manual17Table 3: Certifications and Compliances (Cont.)Installation CategoryDescriptionsTerminals on this product may

Seite 23

WARRANTYTektronix warrants that the products that it manufactures and sells will be free from defectsin materials and workmanship for a period of one

Seite 24

Specifications18P5205 Instructi on ManualTable 4: Typical Electrical Characteristics (Cont.)Overdrive Recovery < 50 ns to 10% of final value after

Seite 25

SpecificationsP5205 Instructi on Manual19Table 5: Typical Mechanical Characteristics (Cont.)Unit Weight (probe only) 11 oz (315 g)Shipping Weight (wit

Seite 26 - 250 mv/div

Specifications20P5205 Instructi on ManualTable 6: Nominal Electrical Characteristics (Cont.)Gain Switchable: 1/50 (“50X”) and 1/500 (“500X”)Audible Ov

Seite 27 - Specificatio ns

P5205 Instructi on Manual21MaintenanceRefer to this section for information about maintaini ng and servi cingthe High Voltage Differential Probe.For t

Seite 28

Maintena nce22P5205 Instructi on ManualTo remove a TEKPROBE interfac e pi n, fi rmly grasp the pointe d tipwith pliers and pull the pin out of the con

Seite 29 - Typical Characteristics

Maintena nceP5205 Instructi on Manual23Removing and Replacing the TEKPROBE Interface CollarTo remove the TEKPROBE interface collar, firmly grasp theco

Seite 30

Maintena nce24P5205 Instructi on ManualRemoving the Compensation Box CoversTo open the compensation box, follow these steps:1. Press the optional rele

Seite 31 - Nominal Characteristics

Maintena nceP5205 Instructi on Manual25TabFigure 10: Replacing the compensation box coverPackaging For Return ShipmentIf possible, use the original pa

Seite 32 - Specifications

Maintena nce26P5205 Instructi on Manual

Seite 33 - Maintenance

The following servicing instructions are for use only byqualified personnel. To avoid injury, do not perform anyservicing other than that st ated in t

Seite 36

P5205 Instructi on Manual27Performan c e Verif icationThe following procedure verifies the warranted electrical characteris-tics of the High Voltage D

Seite 37 - Packaging For Return Shipment

Performance Verificati o n28P5205 Instructi on ManualConstruction of Modified BNC AdapterConstruct the modifie d BNC Adapter from a BNC-male-to-dualbi

Seite 38 - Maintena nce

Performance VerificationP5205 Instructi on Manual29VisePliersModified BNC adapterPlastic cover removedFigure 12: Removing plastic covers from bi nding

Seite 39

Performance Verificati o n30P5205 Instructi on ManualNOTE. These procedures assume that you are using an oscilloscopethat automatically displays the c

Seite 40

Performance VerificationP5205 Instructi on Manual3110. Attach the differential probe input l eads (without attachmentaccessories) by sliding the banan

Seite 41 - Performan c e Verif ication

Performance Verificati o n32P5205 Instructi on Manual15. Set the calibration generator for 10 V output. Set the attenuationbutton on the probe to 50X

Seite 42 - -0035-00 (see

Performance VerificationP5205 Instructi on Manual337. Set the generator to 100 MHz, and set the oscilloscope to5 ns/division. Check for ≥ 2.1 V amplit

Seite 43 - Plastic cover removed

Performance Verificati o n34P5205 Instructi on ManualWARNING. To reduce the risk of electric shock, ensure the gene ratoroutput is disabled before mod

Seite 44 - Differential Gain Accuracy

P5205 Instructi on Manual35Replaceable PartsThis section contains a list of the replaceable m odules for the P5205probe. Use this list to identify and

Seite 45 - Banana plug of probe

P5205 Instructi on ManualiTable of ContentsGeneral Safety Summary v...Service Safety Summary vii...

Seite 46 - Bandwidth

Replaceable Parts36P5205 Instructi on ManualTable 8: Parts List Column DescriptionsColumn Column Name Description1 Figure & Index Number Items in

Seite 47

Replaceable PartsP5205 Instructi on Manual37123456Figure 14: P5205 Probe and replaceable accessories

Seite 48

Replaceable Parts38 P5205 Instructi on ManualP5205 and Replaceable AccessoriesFig. &IndexNumberTektronixPart NumberSerial No.EffectiveSerial No.Di

Seite 49 - Replaceable Parts

Replaceable Parts39P5205 Instructi on Manual123NOTE: Parts illustratedwith dashed lines are notreplaceable45Figure 15: Replaceable parts - compensatio

Seite 50

Replaceable Parts40 P5205 Instructi on Manual1Figure 16: P5205 optional accessoriesOptional AccessoriesFig. &IndexNumberTektronixPart NumberSerial

Seite 51

Replaceable Parts41P5205 Instructi on ManualManufacturers Cross IndexMfr .CodeManufacturer Address City, State, Zip Code18359 PYLON CO. INC. 51 NEWCOM

Seite 52 - 38 P5205 Instructi on Manual

Replaceable Parts42 P5205 Instructi on Manual

Seite 53

Table of C ontentsiiP5205 Instructi on ManualBandwidth 32...DC CMRR 33...R

Seite 54 - Optional Accessories

Table of C ontentsP5205 Instructi on ManualiiiList of FiguresFigure 1: P5205 High-Volt age Differential Probe 2...Figure 2: Safety limits (vol

Seite 55 - Manufacturers Cross Index

Table of C ontentsivP5205 Instructi on ManualList of TablesTable 1: Functional Che ck 8...Table 2: Warranted Electrical Chara

Seite 56 - 42 P5205 Instructi on Manual

P5205 Instructi on ManualvGeneral Safety SummaryReview the following safety precautions to avoid injury and preventdamage to this product or any produ