Using Features for Advanced Applications
3–216
TDS 500C, TDS 600B, & TDS 700C User Manual
To obtain an integral math waveform, do the following steps:
1. Connect the waveform to the desired channel input and select that channel.
2. Adjust the vertical and horizontal scales and trigger the display (or press
AUTOSET).
3. Press MORE Math1, Math2, or Math3 (main) Change Math
waveform definition (side) Single Wfm Math (main).
4. Press Set Single Source to (side). Repeatedly press the same button until the
channel source selected in step 1 appears in the menu label.
5. Press Set Function to (side). Repeatedly press the same button until intg
appears in the menu label.
6. Press OK Create Math Waveform (side) to turn on the integral math
waveform.
You should now have your integral math waveform on screen. See Fig-
ure 3–103. Use the Vertical SCALE and POSITION knobs to size and
position your waveform as you require.
To Create a Integral Math
Waveform