Seitenansicht 125
Syntax and Commands
3-52 AFG3000 Series Arbitrary/Function Generators Reference Manual
[SOURce[1|2]]:FREQuency:MODE
This command sets or queries the frequency sweep state. You can select sine,
square, ramp, or arbitrary waveform for sweep. The arbitrary/function generator
automatically changes to the Continuous mode if any waveform is selected other
than sine, square, ramp, or an arbitrary waveform.
Group
Source
Related Commands
[SOURce[1|2]]:FREQuency[:CW|:FIXed]
[SOURce[1|2]]:FREQuency:CENTer
[SOURce[1|2]]:FREQuency:SPAN
[SOURce[1|2]]:FREQuency:STARt
[SOURce[1|2]]:FREQuency:STOP
Syntax
[SOURce[1|2]]:FREQuency:MODE {CW|FIXed|SWEep}
[SOURce[1|2]]:FREQuency:MODE?
Arguments
CW|FIXed means that the frequency is controlled by the
[SOURce[1|2]]:FREQuency[:CW|:FIXed] command. The sweep is invalid.
SWEep means that the output frequency is controlled by the sweep command set.
The sweep is valid.
Returns
CW|FIXed|SWEep
Examples
SOURce1:FREQuency:MODE SWEep
specifies the sweep command set for controlling the CH 1 output frequency.