Seitenansicht 156
Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-83
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:LOW
This command sets or queries the low level of output amplitude for the specified
channel. If your instrument is a dual-channel model and the
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe] command is set to ON, then the
low level of other channel is also the same value.
Group
Source
Related Commands
[SOURce[1|2]]:VOLTage:CONCurrent[:STATe]
Syntax
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:LOW {<voltage>|MINimum|
MAXimum}
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:LOW? {MINimum|MAXimum}
Arguments
<voltage>::=<NRf>[<units>]
where
<NRf> is the low level of output amplitude.
<units>::=[mV | V]
Returns
<voltage>
Examples
SOURce1:VOLTage:LEVel:IMMediate:LOW -1V
sets the low level of CH 1 output amplitude to –1 V.
Seitenansicht 156
1 2 ... 152 153 154 155 156 157 158 159 160 161 162 ... 229 230

Kommentare zu diesen Handbüchern

Keine Kommentare