Commands Listed in Alphabetical Order
BUS:THReshol
d:CH<x>
Sets or returns the threshold for analog channel <x>, where x is the channel
number. This setting applies to all trigger types that use the channel.
Conditions
This command requires a DPO4AUTO, DPO4EMBD, or DPO4COMP application
module.
Group
Bus
Syntax
BUS:THReshold:CH<x> {ECL|TTL|<NR3>}
BUS:THReshold:CH<x>?
Related Commands
TRIGger:A:LEVel:CH<x>
Arguments
ECL species a preset ECL high level of –1.3V.
TTL species a TTL preset high level of 1.4V.
<NR3> species the threshold level in volts.
BUS:THReshold:D<x>
Sets or returns the threshold for digital channel <x>, where x is the digital channel
number. This will apply to all Search and Trigger Types that use the channel.
Conditions
This command requires a DPO4COMP application module or RS-232C.
Group
Bus
Syntax
BUS:THReshold:D<x> {<NR3>|ECL|TTL}
BUS:THReshold:D<x>?
Related Commands
TRIGger:A:LEVel:D<x>
Arguments
ECL species a preset ECL high level of -1.3V.
TTL species a preset TTL high level of 1.4V.
<NR3> species the threshold level in volts.