Tektronix DPO4000 Bedienungsanleitung Seite 294

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 585
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 293
Commands Listed in Alphabetical Order
Group
Power
Syntax
POWer:REFLevel:PERCent:LO W <NR3>; Ranges={D,0.0,100.0}
POWer:REFLevel:PERCent:LO W?
Arguments
<NR3> is the low value in percentage.
Examples
POWER:REFLEVEL:PERCENT:LO W 15 sets the low reference level to 15% of the
signal amplitude.
POWER:REFLEVEL:PERCENT:LO W? might return
:POWER:REFLEVEL:PERCENT:L OW 10 indicating that the percentage low
reference level is set to 90% of the signal amplitude.
POWer:REFLevel:PERCent:MID<1-3>
Sets or returns the mid reference percent level for waveform measurements. Mid3
is specic to the power application.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:REFLevel:PERCent:MI D<1-3> <NR3>; Ranges={D,0.0,100.0}
POWer:REFLevel:PERCent:MI D<1-3>?
Arguments
<NR3>
is the mid value in percentage.
Examples
POWER:REFLEVEL:PERCENT:MI D1 25 sets the mid reference voltage to 25%.
POWER:REFLEVEL:PERCENT:MI D1? might return
POWER:REFLEVEL:PERCENT:MI D1 50.0000 indicating that the mid reference
voltage level is set to 50%.
POWer:RIPPle (No Query Form)
Does a vertical autoset for ripple measurements or sets the vertical offset to 0.
Conditions
This command requires a DPO4PWR application module.
2-284 MSO4000 and DPO4000 Series Programmer Manual
Seitenansicht 293
1 2 ... 289 290 291 292 293 294 295 296 297 298 299 ... 584 585

Kommentare zu diesen Handbüchern

Keine Kommentare