Commands Listed in Alphabetical Order
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A :RUNT:WHEn
{LESSthan|than|EQual|UNEQu al|OCCURS}
SEARCH:SEARCH<x>:TRIGger:A :RUNT:WHEn?
Arguments
OCCURS argument species a trigger event if a runt of any detectable width occurs.
LESSthan argument sets the oscilloscope to trigger if the a runt pulse is detected
with width less than the time set by the SEARCH:SEARCH<x>:TRIGger:A:
RUNT:WIDth command.
than argument sets the oscilloscope to trigger if the a runt pulse is detected with
width than the time set by the SEARCH:SEARCH<x>:TRIGger:A:RUNT:WIDth
command.
EQUal argument sets the oscilloscope to trigger when the pattern is true for a time
period equal to the time period specied in SEARCH:SEARCH<x>:TRIGger:A:
RUNT:WIDth within a ±5% tolerance.
NOTEQual argument sets the oscilloscope to trigger when the pattern is true for a
time period greater than or less than (but not equal) the time period specied in
SEARCH:SEARCH<x>:TRIGger:A:RUNT:WIDth within a ±5% tolerance.
SEARCH:SEARCH<x>:TRIGger:A:RUNT:WIDth
Sets or returns the width setting for a runt trigger search to determine where to
place a mark. <x> is the search number.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A :RUNT:WIDth <NR3>
SEARCH:SEARCH<x>:TRIGger:A :RUNT:WIDth?
Arguments
<NR3> species the minimum width, in seconds.
SEARCH:SEARCH<x>:TRIGger:A:SETHold:C
LOCk:EDGE
Sets or returns the clock slope setting for a setup/hold trigger search to determine