Commands Listed in Alphabetical Order
Examples
POWER:SOA:LOG
:XMAX 1
sets the XMAX value to 1.0000.
POWER:SOA:LOG:XMAX ? might return :POWER:SOA:LOG :XMAX 1.0000E+3
indicating th
at the XMAX value for log plot type is set to 1.0000E+3.
POWer:SOA:LOG:XMIN
Sets or returns the user XMIN value for use in Log SOA calculations.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:SOA:LOG:XMIN <NR3>
POWer:SOA:LOG:XMIN?
Arguments
<NR3> is the XMIN value used for log SOA calculations.
Examples
POWER:SOA:LOG:XMIN 2 sets the XMIN v alue to 2.0000.
POWER:SOA:LOG:XMIN ? might return :PO WER:SOA:LOG:XMI N 100.0000E-3
indicating that the XMIN value for log plot type is set to 100.0000E-3.
POWer:SOA:LOG:YMAX
Sets or returns the user YMAX value for use in Log SOA calculations.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:SOA:LOG:YMAX <NR3>
POWer:SOA:LOG:YMAX?
Arguments
<NR3> is the YMAX value used for log SOA calculations.